TenneyEnvironmental Test Chambers


Specifications

Tenney Temperature / Humidity Cycling Chambers Specifications

Go Back

Classic Tenney Series

10, 20, and 30 cubic foot models

Overall dimensions in inches(mm) for the 10, 20, and 30 cubic foot models:
Model  Non-
humidified
T10S T10C T20S1.5 T20C-3 T20C-4 T30S T30C
Humidified T10RS T10RC T20RS1.5 T20RC-3 T20RC-4 T30RS T30RC
Workspace W 24(610)

30(762)

36(914)

D 27(686) 32(813) 40(1016)
H 28(711) 36(914) 36(914)
Exterior W 30(762) 41(1041) 46(1168)
D 47(1194) 55(1397) 64(1625)
H 74(1880) 84(2134) 84(2134)

Temperature & Humidity

Low C -40 -73

-40

-73

-73

-40

-73

High C +200 +200

+200

+200

+200

+200

+200
Change Rates in Minutes
Time starting at Ambient to listed temperature. Based on empty chamber.
200°
30 30 60 60 60 70 70
150°
20 20 40 40 40 50 50
93°
15 15 25 25 25 25 25
65°
10 10 15 15 15 15 15
10 5 40 10 5 40 10
-18°
20 10 60 15 10 60 15
-40°
60 20 90 30 20 90 25
-54°
25 40 25 35
-65°
35 50 35 50
Live Load Capacity in Watts (Humidity System Off)
Temperature °C
+10° 600 1275 600 1200 1800 600 1600
-18° 450 1000 450 1000 1500 500 1400
-40° 100 850 100 850 1275 250 1200
-54° 600 700 1050 1000
-65° 400 500 750 700
Utilities, etc
Refrigeration (1)1.5HP (2)1.5HP

(1) 1.5HP

(2) 1.5HP

(2) 2HP

(1) 2HP

(2) 2HP

Heater Capacity

2.5 KW

2.5 KW  (3 KW If 208 Option)

Humidifier KW

1 KW

2 KW

  GPH

.5

1.0

AMPS @ 230V. 1 28

30

32 34 35 34 35
AMPS Fuse 35 40 45
Unit Weight LBS. 850 950 1250 1350 1400 1500 1500

Humidity capability: 20% to 98% RH in the dry bulb range of +20C or +85C as limited by a 3 dew point. Test data based on 24C ambient, sea level, 60 Hz. On 50 Hz or higher than 24C, ambient performance may be reduced. Consult factory regarding any special cooling requirements.

40 and 64 cubic foot models

Overall dimensions in inches (mm) for the 40 and 64 cubic foot models:
Model  Non-
humidified
T40S-2 T40C-4 T40C-6 T64S-7.5 T64C-6 T64C-15
Humidified T40RS-2 T40RC-4 T40RC-6 T64RS-7.5 T64RC-6 T64RC-15
Workspace W 40(1016) 48(1219)
D 44(1118) 48(1219)
H 40(1016) 48(1219)
Exterior W 66(1676) 72(1829)
D 73.5(1867) 81.5(2070)
H 80.25(2038) 91(2311)
Temperature & Humidity
Low -40 -73 -73 -40 -73 -73
High +200 +200 +200 +200 +200 +200
Change Rates in Minutes
Time starting at Ambient to listed temperature. Based on empty chamber.
200°
35 45
150°
25 30
93°
10 15
65°
5 10
5 5
-18°
20 10 10
-40°
120 25 30
-54°
45 40 45 20
-65°
60 50 65 35
Live Load Capacity in Watts (Humidity System Off)
Temperature °C
+10°            
-18° 1000 4500
-40° 100 1000
-54° 1200 2000 1200 4000
-65° 900 1500 800 2500
Utilities, etc
Refrigeration (1) 2 HP (2) 2 HP (2) 3 HP (1) 7.5 HP (2) 3 HP (2) 7.5 HP
Heater Capacity 6 6 6 8 8 8
Humidifier KW 2 2 2 2 2 2
  GPH 1.0 1.0 1.0 1.0 1.0 1.0
AMPS

230V. 1

230V. 1 230V. 1 230V. 1 230V. 1 230V. 1
S & C RC & RS

41

50

50 50 38 46 71 80 57 65 71 80
Fuse

50

60

60

60 60 70 100 110 70 80 100 110
Unit Weight LBS.

2740

3540


Standard Features at a Glance:
S = Standard on all models
H = Units with humidity
T = Temperature only units

Proprietary VersaTenn control system

H

Tenney VaporFlo® humidity system

H

Vapor tight interior made of 100% stainless steel

S

Non-settling, asbestos-free insulation

S

Control tolerance of ±0.3°C and ±2% RH typical
after stabilization

S

Platinum RTD temperature sensor

S

Humidity sensor which requires no wet wick
or water supply

S

Semihermetic refrigeration system

S

Low mass nichrome, open wire heating system

S

NEC wiring compliance

S

Vertical-down recirculating conditioning system

S

Double silicon gaskets on doors

S

Automatic defrost for door gaskets

S

3" Phenolic casters

S

Watlow F4 controller

T


Options:
  10, 20, 30 ft3
  40 & 64 ft3

A = Available Option

IEEE interface

A

A

LinkTenn software for Windows®95 that permits
your computer to control up to 10 chambers

A

A

RS422, 423, 232 or 485 interface assemblies

A

A

6-event relay board @ 1 amp each

A

A

Water demineralizer

A

A

Water reservoir for humidity system (5 gallon)

A

A

Recirculating system for humidity water

A

A

Viewing window, thermally insulated and heated

A

A

Interior lighting

A

A

Shelving, adjustable and removable

A

A

Water-cooled refrigeration system

A

A

Alternate power supply wiring

A

A

Automatic CO2 or LN2 cooling boost system

A

A

Additional ports, gloves, etc.

A

A

Rubber tire casters: 5" or 6" (3" steel are standard)

A

A

Thermocouple or electrical feed through terminals

A

A

Special connectors

A

A

GN2 purge system

A

A

Refrigeration taps and pressure gauges

A

A

Remote console for instrumentation

A

A

Recording instruments

A

A

Redundant thermal protection and alarm system

A

A

External dryer for obtaining humidity as low as 5% (to 20°C)

A

A

Air cooled refrigeration system

A

A

Alternate refrigeration and/or heating systems
for faster temperature change rates or increased
product load handling capabilities.

 

A

 

 